Phone: +86-13707314980

Blog

Kindly reach us to get the fastest response
Home / Blog / Ensuring Accuracy: The Critical Calibration of Your Syringe Pump

Ensuring Accuracy: The Critical Calibration of Your Syringe Pump

Nov 11, 2026

Calibration answers one question: does this pump deliver the volume it says it delivers? On a syringe pump the answer rests on two published figures, and they measure different things. Wincom's SRP-810 is specified within plus or minus 2% mechanically and within plus or minus 3% for rate after correct calibration. That one-point gap is not margin for error; it is the syringe, the clamp and the prime. What follows is the condition set, the test points, the method and the record that make a calibration result worth trusting.

Syringe Pump

Two figures that make two different claims

Mechanical accuracy, within plus or minus 2%, describes the drive: how closely the plunger is moved to the position commanded. It is a property of the pump, built from the motor, the lead screw and the clamp that holds the flange. Rate accuracy, within plus or minus 3% after correct calibration, describes how much fluid actually leaves the syringe over time. It includes the drive, the syringe and the line.

The distinction decides what a failure means. A pump that misses 2% has a mechanical problem to service. A pump that meets 2% and misses 3% is usually being measured with a syringe, a clamp position or a prime the calibration does not represent. Both readings can be true at once, which is why a certificate should say which figure was measured and how.

Before the test: make the conditions repeatable

  • Use the syringe brand, size and material the ward runs. Plunger friction varies between manufacturers, and a calibration on a different syringe describes a device nobody operates.
  • Select the matching syringe size in the menu. The programmable window changes with the syringe, and every test point has to sit inside the window for the size in the clamp.
  • Seat the flange in the clamp and engage the plunger driver without preload. A syringe sitting proud starts the run with slack, and the pump delivers that slack as lost volume.
  • Prime to the tip and let pump and fluid settle at room temperature. A cold syringe and cold fluid change viscosity and the volume you weigh.
  • Level the bench and keep the outlet level with the pump. A tube that siphons adds volume the pump did not push.

Test the range you actually run

A single point at mid-range is a weak calibration. Mechanical slack and start-up error dominate at low rates, and low rates are exactly where neonatal and vasoactive infusions sit. Test the low end of the clinical range, a mid-point, and the top of the range in use, and run each point long enough for start-up to fall out of the average.

Syringe size Programmable rate Step
10 ml 0.1 to 300 ml/h 0.1 ml/h
20 ml 0.1 to 600 ml/h 0.1 ml/h
30 ml 0.1 to 900 ml/h 0.1 ml/h
50 or 60 ml 0.1 to 999.9 ml/h 0.1 ml/h
50 or 60 ml 1000 to 1500 ml/h 1 ml/h
  • On a 10 ml syringe that can mean 0.1, 50 and 300 ml/h; on a 50 ml syringe, 0.1, 500 and 999.9 ml/h.
  • Record the duration of each point. A two-minute run at 0.1 ml/h measures almost nothing.
  • Mind the step size. Between 0.1 and 999.9 ml/h the pump resolves to 0.1 ml/h; from 1000 to 1500 ml/h it resolves to 1 ml/h, so a tight tolerance applied at 1200 ml/h is applied to a coarse step.

Low-rate work also rests on concentration standardisation, which Wincom's note on neonatal continuous infusion concentrations works through in detail.

Know what your method can see

Gravimetric testing weighs what came out and converts mass to volume using fluid density. It sees the whole chain: drive, syringe, clamp, line. Travel measurement watches the plunger move and converts distance to volume. It sees the drive only.

That difference matters more than the test equipment. The mechanical figure describes travel, so a travel-based check confirms 2% and says nothing about the 3% delivery figure. A department that calibrates by travel alone certifies the part that rarely fails. If the method is gravimetric, record balance resolution, density and evaporation control, because those three set the uncertainty of the result.

Pass, fail, and the record that proves it

Declare the tolerance before the run, not after it. Then handle a failure in order: repeat with a new syringe of the same brand, re-check clamp and plunger seat and look for a partial occlusion or a compliant tube, and if two runs fail the same point, take the pump out of service with the reading, the syringe lot and the conditions written down.

The record is the deliverable. It should carry the pump serial number, the method, the syringe specification, the test points, the measured volumes, the tolerance applied, the reference used for the balance or weights, the technician, the date and the next due date. A certificate stating a pass without test points cannot be audited or repeated.

What calibration does not cover

  • Alarms. Occlusion, air-in-line, battery and end-of-infusion need a functional test of their own. A pump can calibrate cleanly and still fail to alarm.
  • The drug library and the records. On the SRP-810D these are configuration, not calibration, and they need a version check against the formulary.
  • Battery capacity and mains changeover, which are proven by running the pump, not by weighing fluid.
  • Wear that has not yet crossed the tolerance. A trend from +0.5% to +2.5% across three services is a warning while every result still passes.

The syringe pump built for this test regime

Repeatable results need a device that accepts the syringes you use, states its limits per size and holds the flange the same way every time. Inside the syringe range the choice comes down to rate envelope and audit features, and the SRP-900B is the model to weigh against the SRP-810 before you specify.

When the therapy needs more volume than a syringe holds, the question changes from accuracy per millilitre to continuous delivery: a volumetric pump runs 1 to 1200 ml/h, and the infusion and syringe pump range puts both classes side by side.

Syringe Pump

⭐ Recommended Syringe Pump | Wincom SRP-810

Wincom's Syringe Pump SRP-810 is positioned for wards, theatres and ICUs delivering small-volume and low-rate infusion, with size-specific rate limits running from 0.1 to 300 ml/h on a 10 ml syringe up to 0.1 to 999.9 ml/h on a 50 or 60 ml syringe in 0.1 ml/h steps, rate accuracy within plus or minus 3% after correct calibration, mechanical accuracy within plus or minus 2%, and a voice alarm system. This makes it useful where accuracy has to be demonstrated rather than assumed, because the two figures separate a drive fault from a syringe or setup effect. Where the calibration record has to sit beside an audit trail, the SRP-810D adds a drug library and storage for 2,000 injection records.

Calibration only holds when the pump, the syringe and the record agree.

Send the syringe brands and sizes your wards use, the rate range and the tolerance your quality system applies. Wincom will confirm which SRP-810 limits apply per size and what documentation ships with the unit.

Contact Us

FAQs

What is the difference between mechanical accuracy and rate accuracy?

Mechanical accuracy covers the drive and is stated within plus or minus 2%. Rate accuracy covers delivered volume and is stated within plus or minus 3% after correct calibration. The gap belongs to the syringe, the clamp and the prime.

How often should a syringe pump be calibrated?

The specification does not set an interval. Local policy and the supplier's recommendation do, so confirm both in writing, and shorten the interval after any failed check or a change of syringe brand.

Can I calibrate with whatever syringe is on the shelf?

Only if that syringe is the one in clinical use. Plunger friction and flange dimensions differ between brands, so a result produced on a different syringe cannot be transferred to the ward.

Does a passing calibration mean the pump is ready to use?

No. Calibration covers delivery accuracy only. Alarm function, battery behaviour and the drug library content need separate checks before the pump returns to a patient.

Related product categories

Please Leave Us A Message If You Have Any Questions
Contact Us

E-mail: [email protected]

Tel: +86-731-84176622
+86-731-84136655

Address: Rm.1507,Xinsancheng Plaza. No.58, Renmin Road(E),Changsha,Hunan,China

Phone: +86-13707314980
Copyright © Wincom Medlab Co., Ltd. All Rights Reserved
InquiryInquiryEmailEmailWhatsAppWhatsAppWechatWechat
Wechat